The
Piecal 334Plus is a 4-20mA Automated Loop Calibrator that builds on the Piecal 334 with advanced features unique to PIE calibrators. Easily find current leaks to ground, deteriorating & damp insulation or corroded terminals with patented loop diagnostic technology (US Patent # 7,248,058). Automated stepping (2, 3 or 5 steps) and ramping lets you set the 334Plus at one instrument and check any other point in the loop.
Get better calibrations with a five digit display to 20.000 mA. You can read the display in dim conditions with the built-in backlight. Additional improvements over 334s from other manufacturers include better long term stability, accuracy, reliability & ruggedness plus improved battery life.
PLUS the 334Plus exceeds the performance of many higher cost calibrators to meet today's most demanding requirements while maintaining the familiar, time proven reliability and ease of use of similar calibrators. The PIECAL 334Plus is the new milliamp calibrator standard!
PIECAL Calibrators Competitive Edge
PIECAL Calibrators are
designed and built by members of the same team that designed and built
the calibrators manufactured by Fluke under the Altek label.
To learn more about these competitive differences, download the
PIECAL 334/334Plus vs. Altek 334A/334 sheet.
| General |
Operating Temperature Range
| -20 to 60°C (-5 to 140F)
|
Storage Temperature Range
| -30 to 60°C (-22 to 140F)
|
Relative Humidity Range
| 10 % ≤RH ≤90% (0 to 35C), Non-condensing
|
10% ≤RH ≤70% (35 to 60C), Non-condensing
|
Size (L x W x H)
| 5.63 x 3.00 x 1.60 in (143 x 76 x 41mm)
|
| Weight | 12.1 oz, 0.34 kg (including boot & batteries)
|
| Batteries | Four "AA" Alkaline 1.5V (LR6)
|
Optional AC Adaptors
| 120 VAC 50/60 Hz 240 VAC 50/60Hz
|
Optional NiMh Rechargeable battery kit
| 120VAC for North America Only; charger, four NiMh batteries, AC & DC cords
|
Lower Battery
| Low battery indication with nominal hour of operation left
|
Protection against misconnection
| Over-voltage protection to 135 vrms (rated for 30 seconds)
|
| Display | High contrast graphic liquid crystal display with 0.315" (8.0mm) high digits.
|
Read mA
|
Ranges and Resolution
| 0.00 to 52.000 mA or -25.00 to 300.00% of 4.-20 mA
|
Accuracy Below 24.010 mA Above 24.010 mA
| ≤ ± (0.05% of 24.000mA +0.01 mA) ≤ ± (0.05% of 24.000mA +0.01 mA)
|
| Voltage burden | ≤ 2V at 50 mA |
Overload/Current limit protection | 54 mA nominal |
| Battery life | ≥ 125 Hours nominal ≥ 100 hrs with backlight on
|
| Source/Power & Measure Two Wire Transmitters |
Ranges and Resolution
| 0.000 to 24.000 mA or -25.00 to 125% of 4-20mA
|
Accuracy EZ-Checks at 4 & 20 mA1 0.000 to 24.000 mA
| ≤ ± (0.01 % of Span at 4 mA & 20 mA (±0.005mA) ≤ ± (0.05 % of 24 mA Span (±0.01mA)
|
| Noise | ≤ ± 1/2 Least Significant Digit
|
Temperature effect
| ≤ ± 0.005%/°C of FS
|
| Loop compliance voltage | ≥ 24 DCV @ 20.00mA
|
Loop drive capability - Leak Detection Off
Loop drive capability - Leak Detection On
| 1200 Ω at 20 mA for 15 hours nominal; 950 Ω with Hart Resistor enabled
1000 Ω at 20 mA for 15 hours nominal;
750 Ω with Hart Resistor enabled
|
Battery life
| ≥30 hours at 12 mA nominal; ≥ hours with backlight on
|
1These are internal calibrated cardinal reference points and accuracy is not defined or limited by display resolution
| 2-Wire Transmitter Simulation |
| Accuracy | Same as Source/Power & Measure
|
| Voltage burden | ≤ 2V at 20 mA |
Overload/Current limit protection | 24 mA nominal |
| Loop voltage limits | 2 to 100 VDC (fuse-less protected from reverse polarity connections) |
| Battery life | ≥ 125 hours nominal; ≥ 100 hours with backlight on
|
Voltage Read
|
Range and Resolution
| -99.99 to +99.99 VDC Full Span (FS)
|
| Accuracy | ≤ ± 0.05% of FS
|
Temperature effect
| ≤ ± 100 ppm/°C of FS
|
Input resistance
| ≥ 2 MΩ |
Battery life
| ≥ 125 hours nominal; ≥ 100 hours with backlight on
|